Constant climate testing

KBF series (4.1 cu.ft. / 115 l)

  • Temperature-Humidity-Simulation
  • Temperature range (°C/°F) 0 - 70 / 32 - 158
  • Humidity range: 10%-80% r.H.

Proven superior: Constant climate chambers

  • Proven precision
    • Double-sided horizontal air flow for maximum parameter uniformity
    • Automatic water and wastewater management
    • Responsive vapor pressure humidification
    • Powerful cooling system for safe operation up to 32 °C ambient temperature
    • Stability tests according to ICH guideline Q1A(R2)
  • Proven documentation
    • Documentation in support of GMP/GLP compliance
    • FDA-compliant software APT-COM™
    • IQ, OQ, PQ validation documentation

The long-term constant temperature and humidity values of BINDER constant climate chambers are impressive. The KBF series is ideal for norm-compliant work according to ICH guidelines. As an inexpensive alternative to using multiple types of climate chambers, the BINDER constant climate chambers can be used for a number of applications.

Areas of Application
  • Basic Research
  • Cosmetics Industry
  • Pharmaceutical Industry
  • Food & Beverage
  • Electronically controlled APT.line™ preheating chamber and refrigerating system assuring temperature accuracy and reproducible results
  • Temperature range 0 °C to 70 °C (32 °F to 158 °F) (without humidity)
  • Temperature range 10 °C to 70 °C (50 °F to 158 °F) (with humidity)
  • Humidity range 10 % to 80 % RH
  • MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
    • User friendly LCD screen
    • Easy-to-read menu guide
    • Integrated electronic chart recorder
    • Variety of options for the graphic display of process parameters
    • Real time clock
  • Electronically controlled humidification and dehumidification system with capacitive humidity sensor
  • Suitable for stability tests according to ICH guideline Q1A (R2)
  • Inner glass door with sealing
  • Independent adjustable temperature safety device class 3.1 (DIN 12880), with visual and audible temperature alarm
  • Access port with silicone plug Ø 30 mm (1.18 inch), left side
  • Safety connection kit for water supply and drainage, including water hose up to 1 m (3.3 ft.) Height
  • Ethernet interface for GLP/GMP and the FDA guideline 21 CFR Part 11 compliant APT-COM™ DataControl-System software
  • 2 stainless steel racks
  • BINDER technical specification conformance certificate
  KBF 115 (E5.2)
Exterior dimensions
Width (mm/inch) 885 / 34.8
Height (incl. feet) (mm/inch) 1050 / 41.3
Depth (incl. door handle, I-triangle, connection 80 mm (3.15 inch)) (mm/inch) 730
Wall clearance, rear (mm/inch) 100 / 3.9
Wall clearance, side (mm/inch) 100 / 3.9
Steam space volume (l/cu.ft.) 156 / 5.5
Number of doors (ea.) 1
Inner glass door(s) (ea.) 1
 
Interior dimensions
Width (mm/inch) 600 / 23.6
Height (mm/inch) 483 / 19.0
Depth (mm/inch) 351 / 13.8
Interior volume (l/cu.ft.) 102 / 3.6
Racks (number standard/max.) 2 / 5
Load per rack (kg/lbs.) 30 / 66.1
Permitted total load (kg/lbs.) 100 / 220.5
Weight (empty) (kg/lbs.) 129 / 284.4
 
Temperature data (without humidity)
Temperature range (°C/°F) 0 - 70 / 32 - 158
Temperature variation at 25 °C (77 °F) (± K) 0.2
Temperature variation at 40 °C (104 °F) (± K) 0.2
Temperature fluctuation (± K) 0.1
Max. heat compensation up to 40 °C (104 °F) (W) 200
 
Climate data (with humidity)
Temperature range (°C/°F) 10 - 70 / 50 - 158
Temperature variation
at 25 °C (77 °F) / 60 % RH (± K) 0.2
at 40 °C (104 °F) / 75 % RH (± K) 0.2
Temperature fluctuation
at 25 °C (77 °F) / 60 % RH (± K) 0.1
at 40 °C (104 °F) / 75 % RH (± K) 0.1
Humidity range (% RH) 10 - 80
Humidity fluctuation
at 25 °C (77 °F) / 60 % RH (± % RH) ≤ 2
at 40 °C (104 °F) / 75 % RH (± % RH) ≤ 2
Recovery time after door was opened for 30 sec
at 25 °C (77 °F) / 60 % RH (min.) 15
at 40 °C (104 °F) / 75 % RH (min.) 8
 
Electrical data
Voltage (± 10 %) 50 / 60 Hz (V) 200 - 240, 1 N ~
Nominal power (kW) 2.0
Energy consumption 40 °C (104 °F) / 75 % RH (W) 1) 470
Noise level approx. (dB (A)) 52

1) These values can be used for dimensioning air condition systems.

All technical data are specified for units with standard equipment at an ambient temperature of 25 °C (77 °F) and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time.

 

  • BINDER Pure Aqua Service

    This convenient and flexible water treatment system extends maintenance intervals and is easily implemented independent of water quality.
    The specific advantage: Point-of-use system with water quality display and replaceable filter cartridge 

  • BINDER Data Logger Kits

    The new BINDER Data Logger Kits – Makes independent recording of temperature and humidity data in the BINDER device possible. The tailored product solution contains helpful accessories: from mounting the logger to the BINDER device to cable access assistance to the sensor mount

  • External water supply set

    External water supply set from fresh and waste water container, wiring and pump 

  • APT-COM™ DataControlSystem GLP Edition

    Software for GLP-compliant control, programming and documentation. Allows networking of up to 30 devices or controllers. Meets FDA 21 CFR Part 11 requirements 

  • Access port

    With silicone plugs for introducing external measuring instruments into the chamber, access ports with 30, 50 or 100 mm (1.18, 1.97, 3.94 inch) diameters

  • Calibration certificate & validation

    BINDER can significantly reduce the workload in qualifying and validating devices. Nobody knows our devices as well and has as much experience in certifications as we do 

  • Temperature measurement of the specimen

    Additional PT 100 temperature sensor for accurate temperature measurement of the specimen and digital display of measured values. Recording of measured data via Ethernet or RS 422 interface optional

KBF 115 (E5.2)
Access ports with silicone plug 30, 50, 100 mm (1.18, 1.97, 3,94 inch) O
Securing elements for additional fastening of racks (1 set of 4 pieces) O
Keyboard lock O
Additional PT 100 temperature sensor, flexibly installed, with external connection, including DIN connector (6-pin) O
RS 422 interface for communication software APT-COM™ DataControlSystem O
External fresh water supply set consits of fresh and waste water tank with 20 l (0.7 cu.ft.) each, cabling and pump O
BINDER PURE AQUA SERVICE consisting of disposible cartridge, hose set and measuring unit O
Disposable cartridge for BINDER PURE AQUA SERVICE O
Factory calibration certificate for temperature and humidity. Measurement in center of chamber at 25 °C ( 77 °F) / 60 % RH or at specified values O
Extension to factory calibration certificate for temperature and humidity. Each additional measurement at an additional measuring point or set of values O
Temperature precision measurement according to DIN 12880 and 9-point humidity measurement / factory standard with measurement log and certificate, measured at (25 °C /77 °F) / 60 % RH or at specified values O
External chart recorder for temperature and humidity O
Data Logger Kit TH 70: For the continuous temperature and humidity recording of -40 °C (-40 °F) to 70 °C (158 °F) / 0 % to 100 % RH. Kit includes 1 data logger, 1 attachable combined humidity/temperature sensor with 2 m extension cable and 1 fixture for the connection at the BINDER unit O
Data Logger Kit TH 70/70: With two attachable combined sensors. One for the continuous temperature and humidity recording of -40 °C (-40 °F) to 70 °C (158 °F) / 0 % to 100 % RH. Second one at the data logger for recording the environmental conditions. Kit includes 1 data logger, 2 attachable combined humidity/temperature sensors with 2 m extension cable and 1 fixture for the connection at the BINDER unit O
Data Logger Kit T 220: For the continuous temperature recording of -90 °C (-130 °F) to 220 °C (428 °F). Kit includes 1 data logger, PT 100 sensor with 2 m Teflon extension cable and 1 fixture for the connection at the BINDER unit O
Data Logger converter cable RS 232 to USB 2.0 O
Data Logger Software: Configuration und evaluation software for all BINDER Data Logger Kits, incl. data cable (RS 232) O
Rack, stainless steel O
Shelf, perforated, stainless steel O
Temperature safety device, class 3.3 (DIN 12880) with visual alarm O
4-20 mA analog output for temperature and humidity measurements (e.g. chart recorder connection), with 6-pin DIN socket. Outputs are adjusted automatically as the controller is adjusted O
Zero-voltage relay alarm outputs for temperature (±2 °C) and humidity (±5 % RH), accessible via 6-pin DIN socket, with audible signal that can be switched off (maximum power rating 24 V AC/DC, 2.5 A) O
Waterproof interior socket 230 V AC (max. 200 W), IP65 protected, with corresponding plug. Max. allowed operating temperature 50 °C (122 °F) O
Locking door handle with key O
Interior lighting (15 W) O